Description
Provides energy efficient, convenient, safe and reliable performance for optimal storage temperature environments necessary for a wide range of life science, pharmacy, biological, medical, clinical and industrial applications.
Standard Features:
- Digital 4×20 character LCD display microprocessor temperature controller
- Audible and visual Hi/Lo temperature alarm
- Adjustable control range from 5°C above ambient to 70°C on solid door models and 8°C above ambient to 60°C on glass door models
- Programmable ramp and soak
- Remote alarm contacts
- Continuous product temperature display
- 3 sensor system, 1 product,1 air
- Power supply switch
- Door ajar alarm, audible and visual
- 2 to 10 volt DC output for product temperature (selectable to air temperature)
- Password protection – set-points and factory settings
- 100 event alarm logging date/time display
- Integrated Hi/Lo product temperature alarm test
- Digital calibration air and product
- Real time clock with date and time display
- Door opening counter with 24 hour automatic reset
- Adjustable alarm beep or constant on (key pad selectable)
- Alarm volume adjustable manually
- Low battery test switch with buzzer
- Perimeter anti-condensate door frame (temperature selectable heater wire)
- Glass anti-condensate heater, temperature programmable (glass door models only)
- Sensor failure alarm
- Power failure alarm
- Exterior cabinet front, sides and back are painted white
- Exterior cabinet top and bottom are galvanized steel
- Interior cabinet liner is stainless steel
- Cabinet is foamed-in-place with CFC Free high density polyurethane foam insulation
- Full size door
- Heavy duty door pivot hinges
- Pull door handles
- Magnetic door gaskets
- Centered key door lock(s) (2 keys)
- Four casters (2 locking), 2 leveling legs
- 3 solid stainless shelves
- Humidity pan – 9 quarts
- CO2 sample port
- Water condensation drain (external) with plug and tube
- Top mounted forced air heating system
- Ambient operating conditions: 15°C to 40°C, Humidity 80% RH maximum non condensing
- Voltage tolerance +/- 10% rated line voltage
- UL/CUL listed
Product Literature: large-capacity-co2-incubators